Basic Info.

Model NO.
BTCY-5280
Certification
CE
Warranty
1 Year
Hatching Rate
98%
Use Voltage
220V-240V
Material
Special Plastic
Package
Wooden Case
Transport Package
Sea Ship, Air Ship
Specification
Medium
Trademark
BOTO
Origin
China
Production Capacity
5000 Months/Mouth

Product Description

 

Xenon Accelerated Aging Test Chamber for Electronic Laboratory

 

Product Description

 

Machine Summary:

Xenon Accelerated Aging Test Chamber is a kind of comprehensive climate test equipment, support climate

aging test,polymer light fastness test(Polymer materials exposed to simulated solar spectrum of artificial

lightthrough the glass to assess the light resistance of materials). From the light, temperature, rain

orcondensation, humidity and so on to do the simulation and strengthen test; use simulating full sunlight

spectrum xenon arc lamp to simulate the destructive waves of different environment. Can be used for

theselection of new materials, improvement of existing materials or evaluate the material durability

Aftermaterial composition changes and other testing.

The Xenon lamp Simulate the uv light, and the machine make the test specimen expensed to uv light,high temperature and water spray to proceed the test! This machine use xenon lamp tube to simulate thesunlight

radiation, use the condensate and spray to simulate the dew and rain.

This machine only need a few days or weeks time, can show the condition or the parts components of aging damage status outdoor many months even a few years time, including colour fading, color change, fracture,

fissure, wrinkle, blister, embrittlement, strength reduction, oxidation and so on.

Electronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test Chamber

Technical Parameters:

Model BT- Z-2025 Xenon Accelerated Aging Test Chamber
Sample exposure area 560cm2
performance index

blackboard or black

mark temperature

BST 35°C-100°C
spectral wavelength UV-A340, wave length:340nm
Irradiation light source

Full spectrum xenon lamp(air-cooled)

irradiance/(w/m2*nm):adjustable

Xenon lamp power 2.5kw (lifetime:2000 hours)

 

Temp. and humi.

control system

time controller

Import programmable time computer integrated

controller(MOELLER)

Accuracy range

Setpoint accuracy: temperature±0.1°C

Indication accuracy: temperature±0.1°C

Temp. Humi. Sensor Platinum resistance Pt100Ω/mv
Heating system

The independent system,Nickel-chromium alloy

electric heating heater

temperature controller

Import integrated microcomputer temperature and

humidity controller

Circulatory system

Low noise of heat-resistant air-condition type motor,

leafy typecentrifugal rotor, Cooling tubes used air-cooled system

Power voltage AC220V±10% 50±0.5HZ
Dimension 81x35x75CM L*D*H

Electronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test Chamber

Maintenance of equipment:

Installation site should consider factors such as the cooling efficiency of the machine and easy to check the maintenance, Other installation considerations:

1.The distance between this machine with wall and any other machines should be more than 60cm.

2.Should be set on the ground of the plain, no vibration, and use gradienter examination.

3.Ambient temperature should be keep into 10°C~25°C, the machine can operate in the most stable status.

If ambient temperature change too fast, for example change 5°C or more in a few minutes, the temperature

and humidity control and cooling rate, will can't very stable control.

4.Should be away from the heat source and inflammable, explosive substance.

5.Do not under direct sunlight, and maintain indoor airiness.

6.Avoid placed in the place of the dust too much.

7.Power supply line and the drain road, should be shortened as much as possible.

Electronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test ChamberElectronic Laboratory Xenon Accelerated Aging Test Chamber