Customization: | Available |
---|---|
After-sales Service: | 12 Months |
Power Supply: | electronic |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specification
Product model |
BT-MHW-25-S Desktop constant temperature test chamber |
Product Applications |
New energy soft polymer battery and button battery constant temperature test |
Sample limit |
This test equipment is prohibited to be used: Test or storage of flammable, explosive and volatile material samples Testing or storage of corrosive samples Test or storage of strong electromagnetic emission source samples Test or storage of radioactive substance samples Test or storage of highly toxic substance samples Test or storage of samples that may produce the above substances or objects during testing or storage |
Inner volume |
25L |
Inner box size |
W280 mm×D250 mm×H360 mm |
Appearance size |
W360 mm×D450 mm×H470 mm |
Net weight of equipment |
About 40kg |
Test environmental conditions |
Environmental temperature+25ºC,Relative humidity≤85%,No sample in the test chamber (no load) |
Temperature range |
15~60ºC |
Temperature fluctuation |
≤1ºC(No load, stable temperature) |
Temperature deviation |
±2.0ºC(No load, stable temperature) |
Heating time |
25ºC→60ºC ≤30 min |
Cooling time |
25ºC→15ºC ≤60 min |
Communication way |
Ethernet standard interface |
Plug wire |
1.5m 10A three-hole plug cable (with leakage protection) |
Transport Size |
The test box is integral, and the whole transportation Maximum shipping size (without packaging): "Refer to 4.3 Dimensions" |
Conditions of Use:The user guarantees the following conditions (the user is responsible for the installation of equipment power supply lines) |
|
Installation site |
good ventilate No strong vibration around the device No strong electromagnetic field around the equipment No flammable, explosive, corrosive substances and dust around the equipment |
Environmental conditions |
Temperature:5ºC~35ºC;Relative humidity:≤85%; Air pressure:86kPa~106kPa |
Power supply condition |
Voltage AC(220±22)V (50±0.5)Hz Power Single phase + protection ground wire 0.2kW Maximum current 1A |
Other |
Opening the door of the test box during the test will cause temperature fluctuations in the box |
Battery specifications and placement |
|
Specifications |
Coin cell or pouch cell |
Placement |
Four-layer placement (each layer can hold up to 8 coin cell) |